Options
High Precision Quantitative Proteomics Using iTRAQ on an LTQ Orbitrap: A New Mass Spectrometric Method Combining the Benefits of All
ISSN
1535-3893
Date Issued
2009
Author(s)
Koecher, Thomas
Pichler, Peter
Schutzbier, Michael
Stingl, Christoph
Kaul, Axel
Penninger, Josef M.
Mechtler, Karl
DOI
10.1021/pr900451u
Abstract
The development of quantitative techniques in mass spectrometry has generated the ability to systematically monitor protein expression. Isobaric tags for relative and absolute quantification (iTRAQ) have become a widely used tool for the quantification of proteins. However, application of iTRAQ methodology using ion traps and hybrid mass spectrometers containing an ion trap such as the LTQ-Orbitrap was not possible until the development of pulsed Q dissociation (PQD) and higher energy C-trap dissociation (HCD). Both methods allow iTRAQ-based quantification on an LTQ-Orbitrap but are less suited for protein identification at a proteomic scale than the commonly used collisional induced dissociation (CID) fragmentation. We developed an analytical strategy combining the advantages of CID and HCD, allowing sensitive and accurate protein identification and quantitation at the same time. In a direct comparison, the novel method outperformed PQD and HCD regarding its limit of detection, the number of identified peptides and the analytical precision of quantitation. The new method was applied to study changes in protein expression in mouse hearts upon transverse aortic constriction, a model for cardiac stress.
File(s)
No Thumbnail Available
Name
Teucher.pdf
Description
closedAccess
Size
348.43 KB
Checksum (MD5)
eb9edb10ea8c18e115aaaf0b58785100