Dose reduction in full-field digital mammography: an anthropomorphic breast phantom study
ISSN
0007-1285
Date Issued
2003
Author(s)
DOI
10.1259/bjr/67597156
Abstract
The aim of this study was to evaluate the potential for radiation dose reduction by using other beam qualities in full-field digital mammography (FFDM) compared with screen-film mammography (SFM). FFDM was performed using an amorphous silicon detector with a caesium iodide scintillator layer (Senographe 2000D, GE, Milwaukee, USA). SFM was performed using a state-of-the-art conventional system (Senographe DMR, GE, Milwaukee, USA) with a dedicated screen-film combination. An anthropomorphic breast phantom with superimposed microcalcifications (50-200 mum) was used to evaluate the detectability of microcalcifications. Contact mammograms and magnification views (m=1.8) performed with both the digital and the screen-film system were compared. Images were exposed automatically. Molybdenum/Molybdenum (Mo/Mo) anode-filter combination, 28 kVp and 63 mAs were selected by the automatic optimization of parameters (AOP) of the conventional system. This exposure protocol (protocol A) was also used as baseline for the digital system. Dose reduction in digital mammography was achieved by using protocol B with Mo/Rh and 31 kVp and protocol C with Rh/Rh and 32 kVp. The detectability of microcalcifications was assessed by 3 experienced readers with a confidence level ranging from 1 to 5. A receiver operating characteristic (ROC) analysis was performed. In protocol A the area under the ROC-curve (A(z)) for contact views performed by the screen-film system was 0.64 and for those performed with the FFDM system 0.68. The A(z) values were 0.74 in protocol B and 0.65 in protocol C for the digital system. For the conventional and digital magnification views A(z) values were 0.71 and 0.79, respectively. For protocol B the A(z) value was 0.81 and for protocol C it was 0.76. There is no statistically significant difference in the A(z) values for the different protocols in digital mammography and no significant difference from the screen-film system. A potential for dose reduction by using other beam qualities seems to be possible with this digital system.