Near-infrared (NIR) spectroscopy. A new method for arthroscopic evaluation of low grade degenerated cartilage lesions. Results of a pilot study
ISSN
1471-2474
Date Issued
2007
Author(s)
Spahn, Gunter
Plettenberg, Holger
Kahl, Enrico
Mueckley, Thomas
Hofmann, Gunther O.
DOI
10.1186/1471-2474-8-47
Abstract
Background: Arthroscopy is a highly sensitive method of evaluating high-grade cartilage lesions but the detection of low-grade lesions is often is unreliable. Objective measurements are required. A novel NIRS (near-infrared-spectroscopy) device for detection of low-grade cartilage defects was evaluated in a preliminary clinical study. Methods: In 12 patients who had undergone arthroscopy, the cartilage lesions within the medial knee compartment were classified according to the ICRS protocol. With a NIR spectrometer system and an optical probe, similar in design to a hook used for routine arthroscopy, the optical properties of cartilage were measured during arthroscopy. Results: The mean ratio of 2 NIR absorption bands of intact cartilage 3.8 ( range 2.3 to 8.7). was significantly lower than that of cartilage with grade 1 lesions (12.8, range 4.8 to 19.6) and grade 2 lesions (13.4, range 10.4 to 15.4). No differences were observed between grade 1 and grade 2 lesions. Conclusion: NIRS can be used to distinguish between ICRS grade 1 lesions and healthy cartilage during arthroscopic surgeries. The results of this clinical study demonstrate the potential of NIRS to objectify classical arthroscopic grading systems.