Options
Multicenter immunoassay validation of cerebrospinal fluid neurofilament light: a biomarker for neurodegeneration
ISSN
1757-6199
1757-6180
Date Issued
2016
Author(s)
Miller, Anne-Marie
Rutkowska, Aleksandra
Bahl, Justyna M. C.
Herukka, Sanna-Kaisa
Koel-Simmelink, Marleen J. A.
Siloaho, Maritta
Skinningsrud, Anders
Zetterberg, Henrik
Teunissen, Charlotte E.
Lawlor, Brian A.
DOI
10.4155/bio-2016-0114
Abstract
Aim: Neurofilament light (NfL) chain, a putative cerebrospinal fluid biomarker, can support neurodegenerative disease diagnosis and indicate disease severity and prognosis. Universal validation protocols when used to measure biomarkers can reduce pre and analytical laboratory variation, thus increasing end-user confidence in the consistency of validation data across sites. Methodology: Here, a commercially available NfL ELISA (UmanDiagnostics, Ume dagger, Sweden) was validated in a multicentered setting using comprehensive newly developed standard operating procedures. Results: The data showed good assay sensitivity and intra and interassay precision. Interlaboratory precision was, however, suboptimal. Conclusion: The UmanDiagnostics assay is suitable for the quantification of NfL in human cerebrospinal fluid. However, sources of interlaboratory variation in the data require further investigation.