Radiation exposure in full-field digital mammography with a flat-panel x-ray detector based on amorphous silicon in comparison with conventional screen-film mammography.
ISSN
1438-9029
Date Issued
2000
Author(s)
DOI
10.1055/s-2000-8370
Abstract
Purpose: Comparison of radiation exposure between a digital amorphous silicon and a screen-film based mammography system. Evaluation of a possible potential of full-field digital mammography in order to decrease the radiation dose. Methods: The average glandular dose for phantom thicknesses from 30 to 60 mm was calculated from experimentally determined entrance surface air kerma for a digital and a conventional mammography system. The effect of reducing the detector dose and of changing the radiation quality on radiation exposure and on image quality were investigated. Results: By using the delivered settings of the automatic exposure control (AEC) devices, both mammographic systems needed nearly the same doses. Regulations and guidelines on radiation doses were complied. With the digital system, a reduction of radiation exposure of up to 40% by using a higher radiation quality and decreasing slightly the detector dose without loss of diagnostic image quality, might be possible. Conclusion: The potential of fullfield digital mammography for radiation dose reduction, as shown in the present phantom study, needs however, a careful examination under clinical conditions.