Diagnostic performance of automated plasma amyloid-β assays combined with pre-analytical immunoprecipitation
Date Issued
2022-09-07
Author(s)
Vogelgsang, Jonathan
Manuilova, Ekaterina
Bauer, Chris
Jethwa, Alexander
Jahn-Brodmann, Anke
Osterloh, Dirk
Lachmann, Ingolf
Breitling, Benedict
Rauter, Carolin
Hansen, Niels
Palme, Stefan
Schuchhardt, Johannes
DOI
10.1186/s13195-022-01071-y
Abstract
Abstract
Background
Measurements of the amyloid-β (Aβ) 42/40 ratio in blood plasma may support the early diagnosis of Alzheimer’s disease and aid in the selection of suitable participants in clinical trials. Here, we compared the diagnostic performance of fully automated prototype plasma Aβ42/40 assays with and without pre-analytical sample workup by immunoprecipitation.
Methods
A pre-selected clinical sample comprising 42 subjects with normal and 38 subjects with low cerebrospinal fluid (CSF) Aβ42/40 ratios was studied. The plasma Aβ42/40 ratios were determined with fully automated prototype Elecsys® immunoassays (Roche Diagnostics GmbH, Penzberg, Germany) by direct measurements in EDTA plasma or after pre-analytical Aβ immunoprecipitation. The diagnostic performance for the detection of abnormal CSF Aβ42/40 was analyzed by receiver operating characteristic (ROC) analysis. In an additional post hoc analysis, a biomarker-supported clinical diagnosis was used as a second endpoint.
Results
Pre-analytical immunoprecipitation resulted in a significant increase in the area under the ROC curve (AUC) from 0.73 to 0.88 (p = 0.01547) for identifying subjects with abnormal CSF Aβ42/40. A similar improvement in the diagnostic performance by pre-analytical immunoprecipitation was also observed when a biomarker-supported clinical diagnosis was used as a second endpoint (AUC increase from 0.77 to 0.92, p = 0.01576).
Conclusions
Our preliminary observations indicate that pre-analytical Aβ immunoprecipitation can improve the diagnostic performance of plasma Aβ assays for detecting brain amyloid pathology. The findings may aid in the further development of blood-based immunoassays for Alzheimer’s disease ultimately suitable for screening and routine use.
Background
Measurements of the amyloid-β (Aβ) 42/40 ratio in blood plasma may support the early diagnosis of Alzheimer’s disease and aid in the selection of suitable participants in clinical trials. Here, we compared the diagnostic performance of fully automated prototype plasma Aβ42/40 assays with and without pre-analytical sample workup by immunoprecipitation.
Methods
A pre-selected clinical sample comprising 42 subjects with normal and 38 subjects with low cerebrospinal fluid (CSF) Aβ42/40 ratios was studied. The plasma Aβ42/40 ratios were determined with fully automated prototype Elecsys® immunoassays (Roche Diagnostics GmbH, Penzberg, Germany) by direct measurements in EDTA plasma or after pre-analytical Aβ immunoprecipitation. The diagnostic performance for the detection of abnormal CSF Aβ42/40 was analyzed by receiver operating characteristic (ROC) analysis. In an additional post hoc analysis, a biomarker-supported clinical diagnosis was used as a second endpoint.
Results
Pre-analytical immunoprecipitation resulted in a significant increase in the area under the ROC curve (AUC) from 0.73 to 0.88 (p = 0.01547) for identifying subjects with abnormal CSF Aβ42/40. A similar improvement in the diagnostic performance by pre-analytical immunoprecipitation was also observed when a biomarker-supported clinical diagnosis was used as a second endpoint (AUC increase from 0.77 to 0.92, p = 0.01576).
Conclusions
Our preliminary observations indicate that pre-analytical Aβ immunoprecipitation can improve the diagnostic performance of plasma Aβ assays for detecting brain amyloid pathology. The findings may aid in the further development of blood-based immunoassays for Alzheimer’s disease ultimately suitable for screening and routine use.
File(s)
No Thumbnail Available
Name
13195_2022_Article_1071.pdf
Size
2.39 MB
Checksum (MD5)
4b30e69d11091d8b0b903bad497232d2