Multi-center evaluation of analytical performance of the microparticle enzyme immunoassay for sirolimus
ISSN
0009-9120
Date Issued
2006
Author(s)
Wilson, Dawn M.
Johnston, F.
Holt, D.
Moreton, M.
Engelmayer, J.
Gaulier, J. M.
Luthe, Hilmar
Marquet, Pierre
Moscato, D.
Oellerich, M.
Mosso, R.
Brunet, M.
Fillee, C.
Schmid, R.
Wallemacq, P.
Barnes, G.
DOI
10.1016/j.clinbiochem.2006.01.017
Abstract
Objectives: This study evaluated the analytical characteristics of the new Abbott microparticle enzyme immunoassay (MEIA) for sirolimus. Design and methods: The protocol consisted of nine sections: evaluation of antibody specificity, linearity, detection limit, quantification limit, endogenous interferents, exogenous interferents, precision, proficiency testing panel, and method comparison. Results: The mean analytical detection limit was 0.68 mu g/L. The sirolimus concentration corresponding to a total CV of 20% was 1.5 mu g/L. Linearity of response was demonstrated across the dynamic range of the assay. Total precision (CVs) at QC control levels from 5 to 22 mu g/L ranged from 5.7 to 12.6%. Assay standardization was found to be in good agreement with LC/MS/MS as compared with target values for spiked sirolimus proficiency samples from an international sirolimus proficiency testing program. Good correlations (R values) of the immunoassay were observed in comparisons to LC/MS/MS. R values tended to be lower in comparisons with LC/UV methods. Across both LC-based methods and all study sites, there was approximately 25% overall positive slope bias due to cross reactivity of the MEIA antibody to metabolites of sirolimus. The assay cross-reactivity to metabolites of sirolimus parent drug ranged from 6 to 63%. Assay interferences were minimal with the exception of hematocrit, which presented a negative relationship to measured sirolimus concentration. Conclusions: The MEIA demonstrated acceptable analytical characteristics for use for routine monitoring of sirolimus immunosuppressive therapy, and is a viable alternative to HPLC-based methods for sirolimus monitoring. (c) 2006 Published by The Canadian Society of Clinical Chemists.