Preclinical evaluation by flat-panel detector-based volumetric CT versus MRI of intervertebral spacers implanted in a porcine model
ISSN
1529-9430
Date Issued
2007
Author(s)
DOI
10.1016/j.spinee.2006.03.017
Abstract
BACKGROUND CONTEXT: Image quality and implant detectability by conventional imaging methods are suboptimal for perioperative spinal diagnostics, primarily limited by implant-related artifacts. PURPOSE: To evaluate the imaging quality of various intervertebral spacers examined by flat-panel detector-based volumetric computed tomography (FD-VCT) versus magnetic resonance imaging (MRI). STUDY DESIGN/SETTING: A preclinical comparative study on an experimental porcine model. The study was performed at a university research facility. METHODS: Three different intervertebral spacer types (titanium, carbon fiber-reinforced polymer, cobalt-chrome-molybdenum) were implanted in a cadaveric porcine spine and then examined by MRI using T1-weighted spin echo (T1w-SE) and turbo spin echo (T1w-TSE) sequences. Comparative imaging was performed with an experimentally approved FD-VCT prototype featuring two-dimensional and three-dimensional imaging and high isotropic spatial resolution. Data analysis focused on spacer shape, implant positioning, and implant-bone interface. RESULTS: Compared with MRI, and despite the use of T1w-SE and T1w-TSE sequences, the image quality and delectability of all target characteristics were better with FD-VCT absent the usual artifacts. Using its option for implant-specific imaging, the experimental FD-VCT imager allowed reliable determination of additional variables such as dimension and volume. CONCLUSIONS: This experimental study provides initial evidence that FD-VCT produces excellently sharp, high-accuracy, artifact-free imaging quality that is superior to MRI in distinguishing key characteristics of intervertebral implants in a preclinical setting. (c) 2007 Elsevier Inc. All rights reserved.