Options
A STED microscope aligned by design
ISSN
1094-4087
Date Issued
2009
Author(s)
DOI
10.1364/OE.17.016100
Abstract
STED microscopes are commonly built using separate optical paths for the excitation and the STED beam. As a result, the beams must be co-aligned and can be subject to mechanical drift. Here, we present a single-path STED microscope whose beams are aligned by design and hence is insensitive to mechanical drift. The design of a phase plate is described which selectively modulates the STED beam but leaves the excitation beam unaffected. The performance of the single-beam setup is on par with previous dual-beam designs.