Low frequency voltage clamp: recording of voltage transients at constant average command voltage
ISSN
0165-0270
Date Issued
2000
Author(s)
DOI
10.1016/S0165-0270(00)00223-5
Abstract
We implemented a simple feedback system that modifies the conventional current clamp mode of a patch clamp amplifier so that transient potentials, such as action potentials and synaptic potentials, can be measured as in the usual current clamp, while the average membrane potential is kept constant at a value chosen by the user. The circuit thus works like the current clamp for high frequency signals and like a voltage clamp for low frequency signals. We delineate its transfer properties and give application examples. (C) 2000 Elsevier Science B.V. All rights reserved.
File(s)
No Thumbnail Available
Name
1-s2.0-S0165027000002235-main.pdf
Description
Artikel
Size
147.81 KB