A novel knee endoprosthesis with a physiological joint shape Part 1: Biomechanical basics and tribological studies
ISSN
0177-5537
Date Issued
2009
Author(s)
Frosch, Karl-Heinz
Floerkemeier, Thilo
Abicht, C.
Adam, Patrick
Fanghaenel, Jochen
Kubein-Meesenburg, Dietmar
Naegerl, H.
DOI
10.1007/s00113-008-1550-5
Abstract
The natural tibiofemoral joint (TFJ) functions according to a roll-glide mechanism. In the stance phase (0-20A degrees flexion), the femur rolls backwards over the tibia plateau, while further flexion causes increased gliding. This kinematics is based on the principle of a quadruple joint. The four morphological axes of rotation are the midpoints of the curvatures of the medial and lateral femoral condyles and the medial and lateral tibia plateau. In addition, the medial and lateral compartments are shifted a few millimetres in a sagittal direction, the medial tibia plateau being concave and the lateral plateau convex. In most knee arthroplasties, these factors are not taken into account; instead they are equipped with symmetrical medial and lateral joint surfaces. Thereby, the midpoints of the curvatures of the sagittal contours of the lateral and medial joint surfaces, on the femoral as well as on the tibial sides, create a common axis of rotation which does not allow a physiological roll-glide mechanism. The goal of this study was therefore to report on the biomechanical basis of the natural knee and to describe the development of a novel knee endoprosthesis based on a mathematical model. The design of the structurally new knee joint endoprosthesis has, on the lateral side, a convex shape of the tibial joint surface in a sagittal cross section. Furthermore, from a mathematical point of view, this knee endoprosthesis possesses essential kinematic and static properties similar to those of a physiological TFJ. Within the framework of the authorization tests, the endoprosthesis was examined according to ISO/WC 14243 in a knee simulator. The abrasion rates were, thereby, lower than or at least as good as those for conventional endoprostheses. The presented data demonstrate a novel concept in knee arthroplasty, which still has to be clinically confirmed by long term results.