Options
Safety of 5 kHz tACS
ISSN
1876-4754
1935-861X
Date Issued
2014
Author(s)
Saiote, Catarina
Opitz, Alexander
Ambrus, Geza Gergely
DOI
10.1016/j.brs.2013.08.004
Abstract
Background: Sinusoidal transcranial alternating current stimulation (tACS) at 5 kHz applied for 10 min at 1 mA intensity over the hand area of the primary motor cortex (M1) results in sustained changes in cortical excitability as previously demonstrated. Objective: Here we have assessed safety aspects of this stimulation method by measuring neuron-specific enolase (NSE) levels, examining electroencephalogram (EEG) traces and analyzing anatomical data by using magnetic resonance imaging (MRI). Methods: Altogether 18 healthy volunteers participated in the study. tACS was applied at 5 kHz for a duration of 10 min over the left M1 at an intensity of 1 mA. Results: After stimulation no significant changes were detected in NSE levels, no structural alterations were observed in the anatomical scans and no pathological changes were found in the EEG recordings. Conclusions: Our data imply that the application of tACS is safe at least within these parameters and with these applied protocols. (C) 2014 Elsevier Inc. All rights reserved.